Steam Boiler / LOCA Test Facility
Following testing can be taken :
- Severe Accident Qualification Test
- Steam Chamber Test / LOCA TEST ( Loss of Coolant Accident)
- MSLB (Main Steam Line Break
- Tests on Active and Passive Valves
Following parameters can be recorded during the testing :
- Chamber pressure
- Chamber temperature
- Differential Pressure sensor available
- RTD & Data recorder available
- DC Power supply available
Features:
- IBR No - MR 18164
- Steam Output - 100Kg/Hr
- Max. Working Pressure - 15Kg/cm2
- Max. Temperature - 200 ˚C
- Test Vessel (chamber) Size - Ø750 mm x 1675 mm
Feed Water:
- Total Hardness in term of calcium carbonate less than 50 PPM
- Total dissolved solids less than 200PPM
- PH value between 8 to 9
- Water conductivity of 250 Micro mhos/cm